Model 34A


FIGURE 1
(The PicoprobeR Model 34A can be mounted on any
micropositioner for use with any probing system.)
The PICOPROBER MODEL 34A was engineered to meet the stringent demands of advanced high frequency circuit designers. This high impedance probe combines full DC capability, rise/fall times of 120 ps, and a nominal loading input impedance of 10 megohm shunted by 0.1 pf. Signal attenuation is 20:1 with a 50 ohm oscilloscope input. Like our other active probes, the Model 34A achieves its bandwidth using just one probe point, which greatly simplifies internal node testing without sacrificing performance. An assortment of user replaceable probe tips are available for a variety of probing needs. The Model 34A is powered by our standard PicoprobeR power supply.
Specifications  
Input Capacitance 0.1pf  
Input Resistance 10.0 megohm  
Rise/fall time 120 ps (5V pulse)  
Frequency Response DC to 3 GHz (-3dB)  
Operating Range -7V to +10V  
Linearity 0.5%  
Gain Accuracy กำ3%  
Signal Attenuation 20 : 1  
     

FIGURE 2
Model 34A Replacement Tips
PartNumber Tungsten Probing WireShaft Diameter Point Radius  
34A-4-10 10 micron <0.1 micron
34A-4-22 22 micron <1.0 micron
34A-4-35 35 micron <2.0 micron
34A-4-60 60 micron <3.0 micron  
34A-4-125 125 micron <5.0 micron  
       

FIGURE 3 photo not to scale
Requires the use of a PicoprobeR Power Supply. Please specify 110V or 220V when ordering.