Model 18B & 19

Features

¡P High Input Impedance
¡P Low Input Capacitace
¡P DC to 350 MHz
¡P Can be mounted on any micropositioner for use with any probing system

FIGURE 1

High impedance probes, PICOPROBER MODEL 18B and MODEL 19, combine the most advanced MOS and bipolar technologies with special, low capacitance packaging techniques to achieve truly remarkable electronic measurement capabilities. While being manufactured each instrument is individually optimized for the best possible performance. The extremely low input capacitance, high input impedance, and almost negligible input leakage current permits the direct probing of even the most sensitive MOS dynamic nodes. At the same time, the full dc capability of this Picoprobe coupled with the high speed capability permits the full characterization of even the fastest circuits.

The 10 micron tungsten probe wire is tapered to an extremely fine point to allow the probing of lines less than 0.3 micron. The fine probe wire flexes when in contact with the circuit, so that damage to the circuit and probe point is minimized. Also the flexing tends to keep the probe point in contact with the circuit even in the presence of probe table vibrations. Model 18B and Model 19 probe tips are also available with a 20 and 50 micron tungsten probe wire sharpened to less than 1 and 3 microns respectively.

The circuitry located in the Model 18B and Model19 PicoprobeR body is very rugged; however, the unprotected MOS input in each probe tip is subject to destruction by electrostatic discharge. Should the probe tip become damaged, it can easily be removed and replaced.

Specifications Model 18B Model 19
Input Capacitance* 0.02pF 0.02pF
Input Leakage 10 femtoamps 10 femtoamps
Rise/Fall Time 1.2ns 1.2ns
Frequency Response dc to 350 MHz dc to 350 MHz
Operating Range 0 to + 15V range -7 to + 8V
Linearity 0.2% 5V range / 2.0% 15V range 0.2% ¡Ó 3V range / 2.0% -7 to +8V range
Gain Accuracy ¡Ó5% ¡Ó5%
Signal Attenuation (High input impedance oscilloscope) 10 to 1
(50 ohm input) 20 to 1
(High input impedance oscilloscope) 10 to 1
(50ohm input) 20 to 1
     
*For 3ns or longer rise and fall signals. Speed limitations of the capacitance cancelling circuitry results in approximately 0.06pF input capacitance for 1ns or shorter rise or fall inputs.
Requires the use of a Picoprobe Power Supply. Please specify 110V or 220V when ordering.

FIGURE 2
PicoprobeR Model 18B and Model 19 Replacement Tips
Part Number >Tungsten Wire Shaft Diameter Point Diameter  
18B-1-10 10 micron <0.1 micron
18B-1-20 20 micron <1.0 micron
18B-1-50 50 micron <3.0 micron
       
       
18B-2-10 10 micron <0.1 micron
18B-2-20 20 micron <1.0 micron
18B-2-50 50 micron <3.0 micron
     
       
18B-4-10 10 micron <0.1 micron
18B-4-20 20 micron <1.0 micron
18B-4-50 50 micron <3.0 micron
       